An imaging system of X-ray grating interferometry, i.e., Talbot-Lau interferometry, was created with a medical X-ray tube, a multi-slit, two X-ray gratings, and an X-ray detector on the basis of design with wave-optic simulation. Imaging performance of the system was examined so that bone cartilage of a chicken wing was depicted clearly with X-ray dose at 3 mGy. Using this imaging system, a small angle scattering image of a cherry was obtained successfully to depict its vascular bundle of the inner structure of the cherry that was invisible in its X-ray absorption image. This imaging trial suggests that the X-ray Talbot-Lau interferometry is a promising tool to detect soft tissue in human body for X-ray image-diagnosis besides the magnetic resonance imaging.<br>