- TANAKA Tomohiko
- Central Research Laboratory, Hitachi, Ltd.
Bibliographic Information
- Other Title
-
- 超音波Vector Flow Mappingの計測信頼性
- チョウオンパ Vector Flow Mapping ノ ケイソク シンライセイ
Search this article
Abstract
<p>Ultrasonic vector flow mapping (VFM) is a method to visualize cardiac flow using vectors. In this article, the VFM algorithm estimating the blood flow vector based on ultrasound doppler measurement and a two dimensional mass conservation law (2D flow assumption) for cardiac flow is briefly reviewed. Since the VFM accuracy can be easily degraded due to a violation of the 2D flow assumption in real situations, the VFM accuracy was validated with a heart phantom by comparing it with particle image velocimetry (PIV). Moreover, a novel method, called a posteriori “VFM accuracy estimation (VAE)” which can estimate the VFM uncertainty for each patient, is introduced to make it possible to refine the quality of the VFM-based diagnosis.</p>
Journal
-
- Oyo Buturi
-
Oyo Buturi 88 (6), 404-407, 2019-06-10
The Japan Society of Applied Physics
- Tweet
Details 詳細情報について
-
- CRID
- 1390282752327760512
-
- NII Article ID
- 130007709491
-
- NII Book ID
- AN00026679
-
- ISSN
- 21882290
- 03698009
-
- NDL BIB ID
- 029781743
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- CiNii Articles
-
- Abstract License Flag
- Disallowed