生体埋込時を想定した空心偏平型経皮トランス

書誌事項

タイトル別名
  • Air-core Transcutaneous Transformers Embedded in a Biological Tissue:
  • 生体埋込時を想定した空心偏平型経皮トランス : 等価回路の提案・評価と経皮トランスの設計
  • セイタイ ウメコミジ オ ソウテイ シタ クウシン ヘンペイガタ ケイヒ トランス : トウカ カイロ ノ テイアン ・ ヒョウカ ト ケイヒ トランス ノ セッケイ
  • Modeling and Validity Evaluation of Equivalent Circuit and Design of a Transcutaneous Transformer
  • -等価回路の提案・評価と経皮トランスの設計-

この論文をさがす

説明

<p> A transcutaneous energy transmitting system (TETS) enables wireless non-penetrating power supply to ventricular assist device. Design of a transcutaneous transformer embedded in a biological tissue has not been studied. Conduction and displacement currents flowing between transmitting and receiving coils may increase in simulated biological tissue as compared to when a TETS’ transformer is placed in air. Therefore energy transmission efficiency is decreased. In this study, we focused on the impedance between transmitting and receiving coils and proposed an equivalent circuit in simulated biological tissue. To evaluate the validity of the proposed equivalent circuit, we compared the transmission efficiency of the measured values with the values derived from the proposed equivalent circuit. Using the proposed equivalent circuit, these values are in good agreement below 2 MHz. It is confirmed that our proposed equivalent circuit is useful to design a transcutaneous transformer embedded in a biological tissue.</p>

収録刊行物

参考文献 (3)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ