An application of Doehlert matrix optimization on developing of flow injection analysis of neomycin in pharmaceutical samples

書誌事項

タイトル別名
  • Original Paper

この論文をさがす

説明

This work proposes a Flow Injection Analysis (FIA) system to determine neomycin in pharmaceutical formulation, utilizing ninhydrin as chromogenic reagent, with absorbance signal measured at 570 nm. To check the significant flow parameters, a screening with 24 full factorial designs was optimized, and a Doehlert matrix used to obtain the optimum variable values. The method presents two linear responses ranging from 29.8 up to 300 mg L-1 (r2=0.9933) and 800.0 up to 2500 mg L-1 (r2=0.9926). The Limits of Detection and Quantifications were 8.94 mg L-1 and 29.8 mg L-1, respectively. The readings frequency was of 33 reading h-1. The method accuracy was checked through comparison with HPLC method and the results found were similar (t-test, 95% confidence level). The method was applied in pharmaceutical samples with very good performance and the precision (RSD) was always below 8.0%. The reagent and sample consumption was very low, what is in agreement with the green chemistry concept.

収録刊行物

  • FIA研究懇談会会誌

    FIA研究懇談会会誌 31 (2), 107-, 2014

    公益社団法人 日本分析化学会 フローインジェクション分析研究懇談会