- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on June 30, 2025】Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Evaluation of analytical performance of intact PTH assay using AIA-CL2400
-
- OBARA Aimi
- Department of Clinical Laboratory, Toranomon Hospital
-
- HERAI Takashi
- Department of Clinical Laboratory, Toranomon Hospital
-
- KAWASAKI Riichi
- Department of Clinical Laboratory, Toranomon Hospital
-
- ENDO Shigeyuki
- Department of Clinical Laboratory, Toranomon Hospital
-
- KOIKE Yukako
- Department of Clinical Laboratory, Toranomon Hospital
-
- YONEYAMA Akiko
- Department of Clinical Laboratory, Toranomon Hospital
Bibliographic Information
- Other Title
-
- AIA-CL2400によるintact PTH測定試薬の基礎的検討
Search this article
Description
<p>We evaluated the basic performance of chemiluminescent immunoassay-based intact PTH assay, which was newly developed as part of the AIA-CL systems. The coefficients of variation (C.V.) of within-run and between-day precisions of the assay using quality control samples or pooled plasma from patients were 1.9 to 3.7% and 3.8 to 5.1%, respectively. Good linearity was observed from 0 to 5,000 pg/mL. The LoB, LoD, and LoQ were 0.19, 0.36, and 1.92 pg/mL, respectively. No interference by endogenous substances was noted. The relationship between AIA-1800 or cobas® 6000 and this method was good. The stability of serum in storage was poor at 4°C. Intact PTH concentrations were reduced by sample transfer to another container. This assay was only reactive against the 7–84 PTH fragment, and reactivity against the other PTH fragments (1–34, 13–34, 44–68, 39–68, 39–84, and 53–84) was below 0.1%. These results suggest that this method is useful for routine examinations, but caution is needed regarding sample stability and transfer.</p>
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 68 (3), 470-475, 2019-07-25
Japanese Association of Medical Technologists
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390282763131420032
-
- NII Article ID
- 130007684129
-
- ISSN
- 21885346
- 09158669
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed