Basic study of urinary oxalic acid measurement using capillary electrophoresis
-
- TSUBOI Isami
- Special Chemistry Department, BML General Laboratory, BML Inc.
-
- MIZUMURA Chie
- Special Chemistry Section, Special Chemistry Department, BML General Laboratory, BML Inc.
-
- KINOSHITA Mizuki
- Special Chemistry Section, Special Chemistry Department, BML General Laboratory, BML Inc.
-
- MACHIDA Satoshi
- Special Chemistry Section, Special Chemistry Department, BML General Laboratory, BML Inc.
Bibliographic Information
- Other Title
-
- キャピラリー電気泳動法を用いた尿中シュウ酸の測定法の基礎的検討
Search this article
Abstract
<p>We performed basic evaluation studies of the urinary oxalic acid measurement by capillary electrophoresis. This method showed CVs of within-run and between-day precisions of less than 6.0%, and the relative error of dilution linearity was good. The recovery test also showed good results. No effects of hydrochloric acid or ascorbic acid on the urinary oxalic acid measurement were detected. In addition, the correlation of urine samples with high-performance ion chromatography and capillary electrophoresis was good (y = 0.971x + 0.380, r = 0.997, n = 50). This study revealed that this method using capillary electrophoresis is applicable in routine testing.</p>
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 69 (2), 179-183, 2020-04-01
Japanese Association of Medical Technologists