Molecular Mass Estimation a Method Using Spectral Change with the Desolvation Temperature in an Electrospray Ionization

Bibliographic Information

Other Title
  • ESIにおける脱溶媒温度変化に伴うスペクトル変化を利用した分子質量推定法
  • ESI ニ オケル ダツヨウバイ オンド ヘンカ ニ トモナウ スペクトル ヘンカ オ リヨウ シタ ブンシ シツリョウ スイテイホウ

Search this article

Description

<p>Electrospray ionization (ESI), the most general ionization method in LC/MS, is a representative example of soft ionization, and usually does not generate fragment ions. However, in some compounds which have a hydroxyl group, [M+H]+, [M+Na]+, [M+NH4]+, etc. are not generated by positive ions, and dehydration from [M+H]+ and [M+H-H2O]+ are mainly observed. If such ions are observed in the measurement of an unknown sample, it will cause inconvenience when acquiring mass information of the analyte. In this study, in the ESI used for on-line LC/MS, the dehydration ions [M+H-H2O]+ from [M+H]+ are mainly observed at normal desolvation temperatures. This paper reports a method for obtaining mass information of an analyte correctly from changes in mass spectral patterns by lowering the desolvation temperature. The performance was demonstrated using cholesterol, eudesmol and hinesol.</p>

Journal

  • BUNSEKI KAGAKU

    BUNSEKI KAGAKU 69 (3), 115-120, 2020-03-05