ナイロン製食品用器具・容器包装のカプロラクタム試験におけるピーク形状改善のための GC 測定条件の検討

書誌事項

タイトル別名
  • GC method modification to improve the peak shape of caprolactam in the official testing method for food utensil, container, and packaging products made of nylon
  • ナイロンセイ ショクヒンヨウ キグ ・ ヨウキ ホウソウ ノ カプロラクタム シケン ニ オケル ピーク ケイジョウ カイゼン ノ タメ ノ GC ソクテイ ジョウケン ノ ケントウ

この論文をさがす

抄録

To improve the peak shape of caprolactam (CPL) on the chromatogram in the official testing method for the food utensil, container, and packaging products made of nylon, we tried to find the cause and modified the GC analytical conditions. The cause of deterioration of CPL peak shape was seemed to be due to the overload of test solution into the GC glass liner. We changed the GC analytical conditions, such as column size, injection volume of test solution, split ratio, inlet temperature, GC oven temperature condition, liner type, and glass wool volume in a liner, and diluted test solution by ethanol or acetone. The peak shape of CPL was considerably improved when the inlet temperature was set to 280ºC at the highest column heat resistant temperature, in addition, the injection volume was changed to half or the test solution was diluted by ethanol or acetone. Furthermore, the peak shape of heptalactam (HPL) that was added to the test solution as an internal standard showed the same peak shape as CPL on the chromatogram. This result suggested that HPL was thought to be available not only for the original purpose, such as correction of sample injection, but also for the confirmation of peak separation of CPL.