- NOMURA Yusuke
- 国立医薬品食品衛生研究所 医療機器部
-
- MIYAJIMA Atsuko
- 国立医薬品食品衛生研究所 医療機器部
-
- UEMATSU Miyuki
- 国立医薬品食品衛生研究所 医療機器部
-
- HAISHIMA Yuji
- 国立医薬品食品衛生研究所 医療機器部
Bibliographic Information
- Other Title
-
- 再生部品の洗浄工程における清浄性評価の留意点
- サイセイ ブヒン ノ センジョウ コウテイ ニ オケル セイジョウセイ ヒョウカ ノ リュウイテン
Search this article
Description
<p>The reprocessing of single-use devices (SUDs) is that the manufacturers responsibility collects the SUDs which have been used at medical facilities, disassemble, clean, replace the parts, reassemble, and sterilize, and then made commercially available after confirming to be as safe and effective as the original SUDs. The Japanese Ministry of Health, Labor and Welfare (MHLW) announced the establishment of a new system for the reprocessing of SUDs on July 31, 2017, and issued a notice on the Pharmaceutical Affairs Law. The cleanliness for reprocessing SUDs is required to be equivalent to that of the relevant original medical devices. However, since international standards for cleanliness assessment of reprocessing SUDs had not been developed, existing guidelines for reusable medical devices are to be considered as a reference. This section describes the trends in Japan and international standards for the cleanliness assessment of reusable medical devices, as well as the evaluation methods for residual protein and residual endotoxin, which are major markers for cleanliness evaluation.</p>
Journal
-
- Regulatory Science of Medical Products
-
Regulatory Science of Medical Products 11 (3), 195-204, 2021
Society for Regulatory Science of Medical Products
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390289553401264640
-
- NII Article ID
- 130008094861
-
- NII Book ID
- AA12678631
-
- ISSN
- 21890447
- 21857113
-
- NDL BIB ID
- 031760484
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL Search
- CiNii Articles
-
- Abstract License Flag
- Disallowed