- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Automatic Translation feature is available on CiNii Labs
- Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Horizontal flow chambers have a shorter residence time in the chamber than vertical flow chambers
-
- Shimizu Kota
- Department of Clinical Engineering, Fujita Health University Hospital
-
- Kuriyama Naohide
- Department of Anesthesiology and Critical Care Medicine, Fujita Health University School of Medicine
-
- Moriyama Kazuhiro
- Laboratory for Immune Response and Regulatory Medicine, Fujita Health University School of Medicine
-
- Kato Masao
- Department of Clinical Engineering, Fujita Health University Hospital
-
- Nakamura Tomoyuki
- Department of Anesthesiology and Critical Care Medicine, Fujita Health University School of Medicine
-
- Hara Yoshitaka
- Department of Anesthesiology and Critical Care Medicine, Fujita Health University School of Medicine
-
- Komura Hidefumi
- Department of Anesthesiology and Critical Care Medicine, Fujita Health University School of Medicine
-
- Inaguma Daijo
- Department of Nephrology, Fujita Health University Bantane Hospital
-
- Nishida Osamu
- Department of Anesthesiology and Critical Care Medicine, Fujita Health University School of Medicine
Bibliographic Information
- Other Title
-
- 水平流入方式チャンバーは垂直流入方式チャンバーよりも,チャンバー内における滞留時間が短い
- スイヘイ リュウニュウ ホウシキ チャンバー ワ スイチョク リュウニュウ ホウシキ チャンバー ヨリ モ,チャンバー ナイ ニ オケル タイリュウ ジカン ガ ミジカイ
Search this article
Description
<p>Introduction : The venous air trap (V) chamber used in blood purification systems is a possible source of coagulation during blood purification. In this study, the fluid residence time of differently shaped V-chambers was evaluated in simulated experiments. Methods : The Prismaflex and ACH-Σ consoles were evaluated. The time from the injection of 1 mL of dye into the V-chamber to its disappearance was measured three times each, and the average values were compared. Additionally, the flow of fluids in the V-chamber was evaluated visually. Results: The fluid residence time in the V-chamber was significantly shorter in the Prismaflex group (p<0.01). Moreover, visual evaluation showed that the fluid in the Prismaflex V-chamber flowed uniformly as a swirl. In contrast, in the ACH-Σ V-chamber, the fluid flowed unevenly pulsating only at the bottom, and stagnation was observed in the upper layer of the chamber liquid surface. Conclusion : It was suggested that structural differences in the V-chamber may affect the retention in the V-chamber. </p>
Journal
-
- Journal of Japan Society for Blood Purification in Critical Care
-
Journal of Japan Society for Blood Purification in Critical Care 13 (1), 54-58, 2022-12-01