歯科矯正アンカースクリューが口蓋へ植立時に破折し迷入した1例

DOI
  • 山城 貴愛
    医療法人渓仁会手稲渓仁会病院歯科口腔外科
  • 石戸 克尚
    富山赤十字病院歯科口腔外科 富山大学学術研究部医学系歯科口腔外科学講座
  • 沖田 美千子
    医療法人渓仁会手稲渓仁会病院歯科口腔外科
  • 原田 雅史
    医療法人渓仁会手稲渓仁会病院歯科口腔外科
  • 金子 剛
    医療法人渓仁会手稲渓仁会病院歯科口腔外科
  • 針谷 靖史
    医療法人渓仁会手稲渓仁会病院歯科口腔外科

書誌事項

タイトル別名
  • A Case of Breakage of Orthodontic Anchor Mini-screws During Implanting in the Palate

この論文をさがす

抄録

Recently, orthodontic anchor mini-screws have been widely used in orthodontic treatment. Fractures of the implants, soft-tissue infections, and injuries of the roots of adjacent teeth are reported as complications of implantation. We report a case of fracture of an orthodontic anchor mini-screw during implantation in the palate. <br>An orthodontic anchor mini-screw was implanted for the patient at a dental clinic. During the procedure, the mini-screw fractured and remained under the mucosa, so she was referred to our hospital to remove the remaining mini-screw. The fractured mini-screw was impacted under the palatal mucosa. Computed tomography revealed a fractured screw in the midline of the palate. We removed the screw under general anesthesia using an ultrasonic bone cutting device. <br>Anchor screw fracture is considered to be caused by excessive torque at the time of implantation. To avoid this problem, it is essential to form the guide hole in advance and then to implant the anchor screw using a torque screwdriver.