パージ・トラップ-ガスクロマトグラフィー質量分析による水道水中の揮発性有機化合物の分析精度に影響を与える要因の解析

書誌事項

タイトル別名
  • Factor Analysis of Affecting Analytical Accuracy of Volatile Organic Compounds in Drinking Water by Purge and Trap-Gas Chromatography/Mass Spectrometry

抄録

<p>Recovery tests of 25 volatile organic compounds (VOCs) spiked into drinking water were performed under various test conditions to evaluate the effects of cooling experimental apparatuses and solvents and diluting water samples during preparation of calibration standard solutions and water samples on analytical precision for testing the VOCs in drinking water using a purge and trap-gas chromatography/mass spectrometry (PT-GC/MS). As a result, compounds with small Henry's constants (i.e., compounds with high volatility) were more affected in trueness and repeatability when cooling was not applied during the test operations. In addition, when the water sample was diluted, the trueness and repeatability was decreased even if cooling was applied during the test operations. Therefore, it is important to conduct test operations with cooling experimental apparatuses and solvents, and without diluting the sample water to obtain better analytical accuracy in VOC analysis of drinking water samples using PT-GC/MS.</p>

収録刊行物

  • 環境化学

    環境化学 34 (0), 1-8, 2024

    一般社団法人 日本環境化学会

詳細情報 詳細情報について

問題の指摘

ページトップへ