Comparison of Echo-Planar Imaging and Compressed Sensing in the Estimation of Flow Metrics from Aortic 4D Flow MR Imaging: A Healthy Volunteer Study

  • Aono Satoru
    Department of Radiological Technology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Tsuneta Satonori
    Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Nishioka Noriko
    Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Aoike Takuya
    Department of Radiological Technology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Hirayama Hiroyuki
    Department of Radiological Technology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Ishizaka Kinya
    Department of Radiological Technology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Kwon Jihun
    Philips Japan Ltd., Tokyo, Japan
  • Yoneyama Masami
    Philips Japan Ltd., Tokyo, Japan
  • Fujima Noriyuki
    Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Sapporo, Hokkaido, Japan
  • Kudo Kohsuke
    Department of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Hokkaido University, Sapporo, Hokkaido, Japan Global Center for Biomedical Science and Engineering, Faculty of Medicine, Hokkaido University, Sapporo, Hokkaido, Japan

抄録

<p>Purpose: Prolonged scanning of time-resolved 3D phase-contrast MRI (4D flow MRI) limits its routine use in clinical practice. An echo-planar imaging (EPI)-based sequence and compressed sensing can reduce the scan duration. We aimed to determine the impact of EPI for 4D flow MRI on the scan duration, image quality, and quantitative flow metrics.</p><p>Methods: This was a prospective study of 15 healthy volunteers (all male, mean age 33 ± 5 years). Conventional sensitivity encoding (SENSE), EPI with SENSE (EPI), and compressed SENSE (CS) (reduction factors: 6 and 12, respectively) were scanned.</p><p>Scan duration, qualitative indexes of image quality, and quantitative flow parameters of net flow volume, maximum flow velocity, wall shear stress (WSS), and energy loss (EL) in the ascending aorta were assessed. Two-dimensional phase-contrast cine MRI (2D-PC) was considered the gold standard of net flow volume and maximum flow velocity.</p><p>Results: Compared to SENSE, EPI and CS12 shortened scan durations by 71% and 73% (EPI, 4 min 39 sec; CS6, 7 min 29 sec; CS12, 4 min 14 sec; and SENSE, 15 min 51 sec). Visual image quality was significantly better for EPI than for SENSE and CS (P < 0.001). The net flow volumes obtained with SENSE, EPI, and CS12 and those obtained with 2D-PC were correlated well (r = 0.950, 0.871, and 0.850, respectively). However, the maximum velocity obtained with EPI was significantly underestimated (P < 0.010). The average WSS was significantly higher with EPI than with SENSE, CS6, and CS12 (P < 0.001, P = 0.040, and P = 0.012, respectively). The EL was significantly lower with EPI than with CS6 and CS12 (P = 0.002 and P = 0.007, respectively).</p><p>Conclusion: EPI reduced the scan duration, improved visual image quality, and was associated with more accurate net flow volume than CS. However, the flow velocity, WSS, and EL values obtained with EPI and other sequences may not be directly comparable.</p>

収録刊行物

参考文献 (16)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ