Basic and application of system reference derivation in electroencephalography

  • ONOZAWA Yuya
    Department of Clinical Laboratory, Kitasato University Hospital
  • SUZUKI Junko
    Department of Clinical Laboratory, Kitasato University Hospital
  • MUNEKATA Shinichi
    Department of Clinical Laboratory, Kitasato University Hospital
  • KANOH Yuhsaku
    Department of Laboratory Medicine, Kitasato University School of Medicine

Bibliographic Information

Other Title
  • 脳波検査におけるシステムリファレンス誘導の基礎と応用

Search this article

Description

<p>Accurate knowledge of the “system reference” is required to record the electrical activity of the brain by digital electroencephalography. The system reference exists in only digital electroencephalography and must be used. The original data of each electrode unit are obtained by preserving the potential difference between the system reference and each electrode. By using the original data to calculate, one can draw the electroencephalogram of each montage. In addition, it is difficult to detect the failure of the differential amplifier by the monopolar reference derivation; therefore, it is necessary to record the system reference derivation in each patient.</p>