Basic and application of system reference derivation in electroencephalography
-
- ONOZAWA Yuya
- Department of Clinical Laboratory, Kitasato University Hospital
-
- SUZUKI Junko
- Department of Clinical Laboratory, Kitasato University Hospital
-
- MUNEKATA Shinichi
- Department of Clinical Laboratory, Kitasato University Hospital
-
- KANOH Yuhsaku
- Department of Laboratory Medicine, Kitasato University School of Medicine
Bibliographic Information
- Other Title
-
- 脳波検査におけるシステムリファレンス誘導の基礎と応用
Search this article
Description
<p>Accurate knowledge of the “system reference” is required to record the electrical activity of the brain by digital electroencephalography. The system reference exists in only digital electroencephalography and must be used. The original data of each electrode unit are obtained by preserving the potential difference between the system reference and each electrode. By using the original data to calculate, one can draw the electroencephalogram of each montage. In addition, it is difficult to detect the failure of the differential amplifier by the monopolar reference derivation; therefore, it is necessary to record the system reference derivation in each patient.</p>
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 68 (4), 800-805, 2019-10-25
Japanese Association of Medical Technologists
- Tweet
Details 詳細情報について
-
- CRID
- 1390564227327729152
-
- NII Article ID
- 130007733909
-
- ISSN
- 21885346
- 09158669
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed