Response evaluation of Onion-like single Bonner sphere neutron spectrometer using TRUST Eu:LiCAF scintillator

Search this article

Description

We have developed the single sphere neutron spectrometer with five sensitive layers using the TRUST Eu:LiCAF scintillator and the WLSF readout. We evaluated the energy dependence of the detector response at thermal neutron and 24 keV, 144 keV, 565 keV, 5MeV and 14.8 MeV mono-energetic neutron irradiation fields. The experimentally determined neutron sensitivities are consistent with the Monte Carlo calculations. The detector responses for a bare and a heavy-water moderated 252Cf sources also show good agreement with the expected ones using the Monte Carlo calculations.