Precise alignment monitor based on an optical frequency comb (3rd report)

DOI

Bibliographic Information

Other Title
  • 光周波数コムを用いた高精度アライメントモニター(第3報)
  • Calibration for the pulse interval of the optical frequency comb without wringing
  • リンギングを用いない光周波数コムパルス間隔の校正

Abstract

<p>We aim to develop an alignment monitor which consists of a measurement grid of multiple absolute distance interferometers, where each interferometer refers pulse interval of an optical frequency comb compressed by an etalon. The uncertainty in our previous calibration for the compressed pulse interval referring a gauge block was restricted, because the uncertainty caused by the wringing between the optical parallels and the gauge block used as the calibration reference can’t be evaluated. Here, we calibrate the pulse interval by referring the gauge brock without wringing.</p>

Journal

Details 詳細情報について

  • CRID
    1390565134829840000
  • NII Article ID
    130007800852
  • DOI
    10.11522/pscjspe.2019a.0_138
  • Text Lang
    ja
  • Data Source
    • JaLC
    • CiNii Articles
  • Abstract License Flag
    Disallowed

Report a problem

Back to top