- Application of Autoexposure Control Unit to the Infant Chest Radiography
この論文をさがす
説明
We replaced a conventional condenser X-ray apparatus with a new inverter type generator for in pediatric chest radiography. Simultaneously the automatic film density control device was examined. This device was consisted of phototimer and (Hitachi-Auto exposure control unit), whith was used for chest radiography in infants and in children less than 10 years old. A sensitivity field of phototimer detector was reformed to half size smaller than of adult.As a results, we could produce good chest radiography of P-A and lateral view in children and reduce the radiation dose to the patient without retake.