- 新谷 英晴
- 厚生省国立衛生試験所療品部
書誌事項
- タイトル別名
-
- Effect of Suspension, Carrier and Culture Medium on D Value upon Preparation of Biological Indicator (BI)
この論文をさがす
説明
The ISO 11138 series on biological indicator (BI) are insufficient for the concrete description on suspension, carrier and recovery culture medium of microorganisms for preparation of BI. Suspension, carrier and culture medium have a potential to affect D value of BI. Individual effect of suspension, carrier or culture medium on D value of BI was reported, but the combination effect of these factors on D value was not reported yet and the author did not confirm individual effect on D value contrary to the previous papers. The combination effect of salt from suspension when drying and pore of the carrier for entrapping microorgamisms will be speculated to prevent sterilization. D value differed depending on constitute of culture medium, so it will be required to confirm if the constitute of soybean casein digest broth in WD 14161 is appropriate or not to attain the exact D value.
収録刊行物
-
- 医科器械学
-
医科器械学 65 (11), 515-518, 1995-11-01
一般社団法人 日本医療機器学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390569923316275456
-
- NII論文ID
- 110002520249
-
- NII書誌ID
- AN00013123
-
- ISSN
- 18814875
- 0385440X
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可