Accuracy and Etalon Fringe Phenomena in Methane-gas Spectrometry Employing Tunable Diode Lasers
Search this article
Description
A method to monitor atmospheric gas concentration employing a tunable diode laser is proposed. An apparatus was built and experiments were made. The tunable diode laser employed was of'PbSnTe with its wavenumber about 1200 cm(-1) at a temperature of 77 K. Methane concentration in laboratory atmosphere was measured on the experiments. The inquired gas-density is obtained by a hardwired microcomputer incorporating the measured absorption spectrum over a prescribed range instead of an absorption at a single wavenumber. By this, drift type noises are suppressed. After an analysis based on experimental data, an attainable sensitivity for methane was found to be 1 ppb.m. This value may increase if an etalon fringe interference exists. The natural concentration, about 1 ppm, of atmospheric methane was actually measured with a short optical path of only one meter.
Journal
-
- Memoirs of the School of Engineering, Okayama University
-
Memoirs of the School of Engineering, Okayama University 15 (2), 79-92, 1981-03-30
岡山大学工学部
- Tweet
Details 詳細情報について
-
- CRID
- 1390572174510441344
-
- NII Article ID
- 120002307246
-
- NII Book ID
- AA00733903
-
- ISSN
- 04750071
-
- DOI
- 10.18926/15776
-
- Text Lang
- en
-
- Article Type
- departmental bulletin paper
-
- Data Source
-
- JaLC
- IRDB
- CiNii Articles