Numerical Estimation of the Correlation between Temperature Rise due to Medical Metal Plate Implants and Spatial Averaging Dimensions of Local SAR When Exposed to Mobile Phone Frequency Band Radiation

DOI

Bibliographic Information

Other Title
  • 携帯電話周波数帯電波ばく露時の医療用埋込金属プレートに起因した温度上昇と局所SAR空間平均寸法の相関検討

Abstract

The temperature increase in biological tissue when a human body with implanted metal plates was exposed to radio waves was evaluated. The numerical estimation based on the FDTD method was performed to calculate the local SAR enhancement. The temperature distribution in the tissue was estimated by solving the bioheat transfer equation using the obtained SAR as the heat source. As a result, due to the difference between SAR distribution and temperature distribution, 0.1 g averaged SAR is highly correlated with temperature rise.