Simultaneous Determination of Inorganic Anions by Sequential Injection Chromatography System Constructed from a Monolithic Column and a Microfluidic Polymer Chip with an Embedded Ion-Selective Electrode

書誌事項

タイトル別名
  • Original Paper

この論文をさがす

説明

A sequential injection chromatography (SIC) constructed from a short monolithic column and a microfluidic polymer chip with an embedded Cl- ion-selective electrode (Cl--ISE) as a detector was developed for simultaneous determination of common inorganic anions. Under the optimal flow conditions of the SIC system, the determination range for Cl-, NO2-, Br-, NO3- ions was as follows: For Cl- ion in a concentration range from 1.0 x 10-3 to 5.0 x 10-2 mol dm-3, for NO2- ion in the concentration range from 1.0 x 10-3 to 5.0 x 10-2 mol dm-3, for Br- ion in the concentration range from 1.0 x 10-3 to 5.0 x 10-2 mol dm-3, and for NO3- ion in the concentration range from 3.0 x 10-4 to 1.0 x 10-2 mol dm-3, respectively. Limit of detection, defined as S/N=3, was ca. 3.0 x 10-4 mol dm-3 for Cl-, NO2-, Br-, NO3- ions.

収録刊行物

  • FIA研究懇談会会誌

    FIA研究懇談会会誌 29 (2), 89-, 2012

    公益社団法人 日本分析化学会 フローインジェクション分析研究懇談会

詳細情報 詳細情報について

問題の指摘