Evaluation of Low-contrast Detectability of Iterative Reconstruction Algorithm in High-speed CT Technology for Head
-
- Masuda Shota
- Department of Radiological Technology, Keio University Hospital
-
- Sugisawa Koichi
- Department of Radiological Technology, Keio University Hospital
-
- Minamishima Kazuya
- Department of Radiological Technology, Keio University Hospital
-
- Yamazaki Akihisa
- Department of Radiological Technology, Keio University Hospital
-
- Watanabe Toshio
- Department of Radiological Technology, Keio University Hospital
Bibliographic Information
- Other Title
-
- 逐次近似画像再構成法を用いた頭部CT 高速撮影の低コントラスト検出能評価
- 臨床技術 逐次近似画像再構成法を用いた頭部CT高速撮影の低コントラスト検出能評価
- リンショウ ギジュツ チクジ キンジ ガゾウ サイコウセイホウ オ モチイタ トウブ CT コウソク サツエイ ノ テイコントラスト ケンシュツノウ ヒョウカ
Search this article
Description
<p>Purpose: The purpose of this study was to assess the feasibility of high-speed CT technology for head without deterioration of low-contrast detectability using the brain LCD (Canon Medical Systems) of iterative reconstruction. Methods: System performance (SP) function analysis, low-contrast object specific contrast-to-noise ratio (CNRLO) analysis, and visual evaluation using Scheffe’s paired comparison were performed. Additionally, analysis of the correlation of CNRLO and visual scores was performed. SP was performed with the self-made phantom. CNRLO was calculated with the catphan 504 phantom (CTP 515). Visual evaluation was performed using the brain phantom which simulated such as cerebral infarction and investigated on a fivepoint scale. All images were acquired with pitch factor of 0.61 (low pitch) and 1.40 (high pitch). All images were reconstructed with filtered back projection (FBP), brain LCD standard (LCD STD) and strong (LCD STR). Results: SP of brain LCD improved compared with FBP. CNRLO of FBP decreased in high pitch compared with low pitch. CNRLO of brain LCD images acquired by low- and high pitch were improved compared with FBP. Visual scores denoted similar trends to that of CNRLO and there was high correlation with CNRLO. Conclusion: It was suggested that using brain LCD can achieve the high speed CT technology for head without deterioration of low-contrast detectability.</p>
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 75 (3), 254-262, 2019
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390845713059222528
-
- NII Article ID
- 130007617569
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 029607918
-
- PubMed
- 30890673
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL Search
- Crossref
- PubMed
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed