Unprecedented high throughput titration by feedback-based and subsequent fixed triangular wave-controlled flow ratiometry and its application to quantification of Japanese Pharmacopoeia drugs
この論文をさがす
説明
Throughput rate of flow ratiometric titration has further been enhanced by shortening the lag time from the confluence of solutions upstream to the sensing of signal downstream and by optimizing analytical parameters. Feedback-based upward and downward scans of titrand/titrant flow ratio were repeated in order to offset the effect of the lag time and thus to locate the equivalence point. Subsequent faster fixed triangular wave-controlled scans in narrower range further increased the throughput rate. Analytical parameters such as scan rate and scan range were optimized. Maximally, 46.9 titrations/min was realized with reasonable precision (RSD = 1.79%). Applicability of the method to the quantitation of the Japanese Pharmacopoeia drugs (furosemide, isoniazid and prochlorperazine maleate) was verified, where the latter two drugs were determined by nonaqueous titrations.
収録刊行物
-
- FIA研究懇談会会誌
-
FIA研究懇談会会誌 36 (2), 97-, 2019
公益社団法人 日本分析化学会 フローインジェクション分析研究懇談会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390846609815113344
-
- NII論文ID
- 130007814532
- 120006881408
- 40022140101
-
- NII書誌ID
- AA10962565
-
- ISSN
- 24337374
- 0911775X
-
- NDL書誌ID
- 030220255
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- IRDB
- NDLサーチ
- CiNii Articles