Evaluation of LUMIPULSE Presto IL-2R for soluble interleukin-2 receptor assay using LUMIPULSE L2400
-
- ISHIHARA Yuri
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- SHIMOSAKA Hironori
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- KURANO Makoto
- Department of Clinical Laboratory, The University of Tokyo Hospital Department of Clinical Laboratory Medicine, Graduate School of Medicine, The University of Tokyo
-
- SATO Tomoaki
- Department of Clinical Laboratory, The University of Tokyo Hospital
-
- YATOMI Yutaka
- Department of Clinical Laboratory, The University of Tokyo Hospital Department of Clinical Laboratory Medicine, Graduate School of Medicine, The University of Tokyo
-
- YASUDA Naomi
- Department of Clinical Laboratory, The University of Tokyo Hospital
Bibliographic Information
- Other Title
-
- 免疫発光測定装置ルミパルスL2400を用いた可溶性インターロイキン-2レセプター測定試薬「ルミパルスプレストIL-2R」の基礎的性能評価
Search this article
Description
<p>A soluble IL-2 receptor (sIL-2R), the circulating form of the IL-2 receptor, localizes on lymphoid and some cancer cells and is used widely as a biomarker for adult T-cell leukemia and non-Hodgkin lymphoma. In this study, we evaluated the performance of a chemical luminescence test reagent LUMIPULSE Presto IL-2R for the measurement of serum soluble IL-2 receptor using an automated immune chemiluminescent system LUMIPULSE L2400. The coefficients of variation for within-run and between-day reproducibilities ranged from 0.7% to 3.3%. Linearity was observed up to about 80,000 U/mL and the lower limit of detection was 2.79 U/mL. A good correlation was observed between Immulyze IL-2R and LUMIPULSE Presto IL-2R. In conclusion, the analytical performance of LUMIPULSE Presto IL-2R seemed acceptable for routine clinical laboratory tests.</p>
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 69 (3), 340-345, 2020-07-25