Development of Ambient Pressure Hard X-ray Photoelectron Spectroscopy at SPring-8

抄録

An ambient pressure photoelectron spectroscopy measurement that uses with hard X-rays (AP-HAXPES) were conducted at the BL36XU of SPring-8. The AP-HAXPES system with a commercial differential pumping-type spectrometer (R4000 HiPP-2, Scienta Omicron Inc.) was installed in the beamline with the excitation light of 7.94 keV focused to a beam size of 20 μm × 20 μm on the sample position. In this report, we replaced the front cone with our home-made one with an aperture diameter of 30 µm to increase the pressure limit in the AP-HAXPES measurement, although the standard aperture size in the spectrometer is a diameter of 300 μm. In addition, we have adopted the working distance of 60 μm in order not to perturb the gas environment at the sample surface. Using the windowless electron spectrometer system, we have succeeded in measuring X-ray photoelectron spectra under real ambient atmosphere (105 Pa).

収録刊行物

参考文献 (2)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘