Miniaturized Low-Pressure Ion-Exchange Module and Its Application to an Acidic Eluent Generator for Open Tubular Ion Chromatography

  • Watanabe Hiroki
    Graduate School of Pharmaceutical Sciences, Tokushima University
  • Sugiura Jun
    Faculty of Pharmaceutical Sciences, Tokushima University
  • Tanaka Hideji
    Faculty of Pharmaceutical Sciences, Tokushima University Graduate School of Biomedical Sciences, Tokushima University
  • Kubáň Petr
    Department of Bioanalytical Instrumentation, CEITEC Masaryk University
  • Takeuchi Masaki
    Faculty of Pharmaceutical Sciences, Tokushima University Graduate School of Biomedical Sciences, Tokushima University

この論文をさがす

抄録

We describe a miniaturized eluent generator (MEG), which can be integrated into open tubular capillary column ion chromatography (OTCC-IC). The MEG consists of one central eluent channel and two feed acid chambers, with each phase isolated by a cation-exchange membrane (CEM) and an anion-exchange membrane (AEM). Pure water flowed through the central eluent channel while acid solution (HA) was fed to both chambers. Applying the electric field to the MEG, H+ and A- simultaneously and respectively through the CEM and AEM to form HA in the central channel. The MEG successfully produced tartaric acid and 2,6-pyridinedicarboxylic acid as the eluent for cation OTCC-IC. The gradient elution improved the peak resolution and analytical window for five common cations (Na+, NH4+, K+, Ca2+, Mg2+).

収録刊行物

  • FIA研究懇談会会誌

    FIA研究懇談会会誌 38 (1), 11-, 2021

    公益社団法人 日本分析化学会 フローインジェクション分析研究懇談会

詳細情報 詳細情報について

問題の指摘