Fundamental study of glycated hemoglobin analyzer, “The Lab 001”

  • MIYASHITA Daichi
    Department of Clinical Laboratory, National University Corporation Gunma University Hospital
  • NAGAI Natsumi
    Department of Clinical Laboratory, National University Corporation Gunma University Hospital
  • SATO Mika
    Department of Clinical Laboratory, National University Corporation Gunma University Hospital
  • KAMIYAMA Keita
    Department of Clinical Laboratory, National University Corporation Gunma University Hospital
  • NAKAJIMA Kiyomi
    Department of Clinical Laboratory, National University Corporation Gunma University Hospital
  • MURAKAMI Masami
    Department of Laboratory Medicine, National University Corporation Gunma University Graduate School

Bibliographic Information

Other Title
  • グリコヘモグロビン分析装置The Lab 001の基礎的検討
  • グリコヘモグロビン ブンセキ ソウチ The Lab 001 ノ キソテキ ケントウ

Search this article

Abstract

<p>We evaluated the performance of the glycated hemoglobin analyzer The Lab 001 in capillary electrophoresis. The accuracy results obtained were good, and the correlation coefficient between the HPLC analysis and capillary electrophoresis was 0.996. In addition, it was shown that The Lab 001 can separate and detect mutated hemoglobin and hemoglobin F. Although the mutated hemoglobin and hemoglobin F can affect HbA1c measurement, The Lab 001 can be used for measurement without being affected by these types of hemoglobin.</p>