Development of the High Sensitivity Measurement Method for Checking Cleanliness of Medical Instrumensts
-
- Shimizu Masatake
- Surgical Center, Osaka University Hospital
-
- Nakata Seizoh
- Surgical Center, Osaka University Hospital
-
- Fushimi Ryo
- Surgical Center, Osaka University Hospital
-
- Mizutani Ayako
- Surgical Center, Osaka University Hospital
-
- noguchi Satoshi
- Surgical Center, Osaka University Hospital
-
- Monden Morito
- Surgical Center, Osaka University Hospital
-
- Isaka Yasuake
- Inui Medics Co., Ltd.
-
- Uraoka Shinji
- Inui Medics Co., Ltd.
Bibliographic Information
- Other Title
-
- 医療器具に付着した蛋白質の高感度測定法について
Search this article
Description
At present, the amide black 10B dyeing method is mainly used to make a final check for cleanliness of medical instruments after washing. But this method inbolber bisual inspertion for the final check and has problems in that it cannot be used to check an inner part of a tubular instrument that is impossible to view and that the check of contamination tends to be influenced by subjective judgment of an inspector. STo solve these problems, we attempted to quantify residual protein by using the protein dyeing agent Coomassie Brilliant Blue G-250. This agent permits quantifying residual contamination in cleaned medical instruments including the portions that cannot be viewed. Coomassie(R) Protein Assay Reagent Kit(PIERCE) was used as a protein dyeing agent. 0.2M sodium hydroxide liquid extrected protein at 50℃ for 30 minutes. Extraction liquid was used as a sample for fixed quantities. The coefficient of variation (C. V %) of this method ia as good as 2.3 to 3.3%. The pervention to this method by the antiseptic was not usually accepted by operating concentration. When the amount of protein which remains in scissors and suction tubes which were washed by the way washing processes, such asa a bliid coagulation prevention agent and wxistence of immersing washing, dirffer was measured by this method, the protein of high concentration which skipped the washing process was detected. These results showed that this method was very effective in the fixed quantity of the residual protein from the whole instrument.
Journal
-
- JAPANES JOURNAL OF MEDICAL INSTRUMENTATION
-
JAPANES JOURNAL OF MEDICAL INSTRUMENTATION 75 (1), 28-33, 2005-01-01
Japanese Society of Medical Insturmmentation
- Tweet
Details 詳細情報について
-
- CRID
- 1390855533654562304
-
- NII Article ID
- 110002521810
-
- NII Book ID
- AN00013123
-
- ISSN
- 18814875
- 0385440X
- http://id.crossref.org/issn/0385440X
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed