Validation of cough peak flow assessment using a spirometer in healthy young adults

Bibliographic Information

Other Title
  • 健常若年成人におけるスパイロメーターを用いて測定した Cough peak flow の検証
  • A validation study of the agreement with Cough peak flow measurements with a peak flow meter and the appropriateness of cough assessment
  • ―ピークフローメーターでのCough peak flow 測定値との一致性と, 咳嗽評価の適否における検証研究―

Search this article

Description

<p>In this study, we investigated the difference in cough peak flow (CPF) values when measured using an automatic spirometer (CPF(S)) and a peak flow meter (CPF(P)) in 44 healthy young adults. The agreement between values was verified using the BlandAltman analysis. Furthermore, the reproducibility of each CPF value and the validity of the cough assessment method were also determined. CPF(S) values were lower than the CPF(P) values, with a significant fixed error (p<0.001) and proportional error (p<0.001). In addition, the intraclass correlation coefficients were 0.900 and 0.941 for CPF(S) and CPF(P),respectively. There was a significant correlation between each CPF and cough-related indices, and the validity was confirmed. Overall, these findings indicate that the CPF(S) and CPF(P) values differed and that the CPF(S) method was sufficiently suitable for cough strength assessment.</p>