- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on June 30, 2025】Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
IMPROVEMENT OF THE SPATIAL RESOLUTION WITH A DECONVOLUTION METHOD FOR A MULTI-PINHOLE SPECT SYSTEM
Bibliographic Information
- Other Title
-
- デコンボリューション法による静止型マルチピンホールSPECTの空間分解能改善
Search this article
Description
We propose a method to improve the spatial resolution of an image for a static multi-pinhole SPECT system. This system has a advantage of measuring dynamic function of organs, which could not be performed by conventional systems. However, the image quality is sensitive to the pinhole size of a collimator. Especially, in the case of a large pinhole, the widely applied 7-rays method could not sufficiently correct the image resolution. Therefore, we propose a point spread function (PSF) deconvolution method for a multi-pinhole SPECT system. In this paper, we assumed a shift-variant PSF at each pixel of the detector, and performs PSF deconvolution on projection data. The feasibility of this method was investigated with a simulation. Projection data were obtained by a Monte Carlo method. The reconstruction images with our proposed method had significantly higher spatial resolution and reproduced more detail structures compared with those of the 7-rays method. This results indicated the effectiveness of our proposed method.
Journal
-
- 法政大学大学院紀要. 理工学研究科編
-
法政大学大学院紀要. 理工学研究科編 63 1-5, 2022-03-24
法政大学大学院理工学研究科
- Tweet
Details 詳細情報について
-
- CRID
- 1390855965512543872
-
- HANDLE
- 10114/00025372
-
- ISSN
- 24368083
-
- Text Lang
- ja
-
- Article Type
- departmental bulletin paper
-
- Data Source
-
- JaLC
- IRDB
-
- Abstract License Flag
- Allowed