IMPROVEMENT OF THE SPATIAL RESOLUTION WITH A DECONVOLUTION METHOD FOR A MULTI-PINHOLE SPECT SYSTEM

Bibliographic Information

Other Title
  • デコンボリューション法による静止型マルチピンホールSPECTの空間分解能改善

Search this article

Description

We propose a method to improve the spatial resolution of an image for a static multi-pinhole SPECT system. This system has a advantage of measuring dynamic function of organs, which could not be performed by conventional systems. However, the image quality is sensitive to the pinhole size of a collimator. Especially, in the case of a large pinhole, the widely applied 7-rays method could not sufficiently correct the image resolution. Therefore, we propose a point spread function (PSF) deconvolution method for a multi-pinhole SPECT system. In this paper, we assumed a shift-variant PSF at each pixel of the detector, and performs PSF deconvolution on projection data. The feasibility of this method was investigated with a simulation. Projection data were obtained by a Monte Carlo method. The reconstruction images with our proposed method had significantly higher spatial resolution and reproduced more detail structures compared with those of the 7-rays method. This results indicated the effectiveness of our proposed method.

Journal

Details 詳細情報について

  • CRID
    1390855965512543872
  • DOI
    10.15002/00025372
  • HANDLE
    10114/00025372
  • ISSN
    24368083
  • Text Lang
    ja
  • Article Type
    departmental bulletin paper
  • Data Source
    • JaLC
    • IRDB
  • Abstract License Flag
    Allowed

Report a problem