Basic study of solute removability of the newly developed continuous hemofilter “SHG”
-
- Miyasaka Takehiro
- Department of Medical Engineering, Faculty of Health Care Sciences, Himeji Dokkyo University
-
- Ozaki Itsumi
- Department of Medical Engineering, Faculty of Health Care Sciences, Himeji Dokkyo University
-
- Okuda Yukihiro
- Department of Medical Engineering, Faculty of Health Care Sciences, Himeji Dokkyo University
-
- Tabuchi Akinari
- Department of Medical Engineering, Faculty of Health Care Sciences, Himeji Dokkyo University
Bibliographic Information
- Other Title
-
- 新規持続緩除式血液濾過器SHGの溶質除去能に関する基礎的検討
Description
<p>A new type of continuous hemofilter, the “Hemofeel SHG” (SHG), was developed by Toray Industries, Inc., Tokyo, Japan. The SHG has a smoother inner surface of hollow fibers for blood flow than the existing “Hemofeel SH” (SH). In this study, we examined filter lifetime and solute removability using bovine blood. Both SHG and SH showed gradual increases in blood inlet pressure. The urea clearance of SHG did not differ significantly from that of SH at 20 min after starting the pump, but was significantly higher when the blood inlet pressure reached 250mmHg (a hemofiltor exchange criterion). In conclusion, based on urea clearance, the lifetime solute removability of SHG can be expected to be prolonged compared to that of SH.</p>
Journal
-
- Journal of Japan Society for Blood Purification in Critical Care
-
Journal of Japan Society for Blood Purification in Critical Care 2 (2), 198-201, 2011-12-01
Japan Society for Blood Purification in Critical Care
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390858220181644544
-
- ISSN
- 2434219X
- 21851085
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
-
- Abstract License Flag
- Disallowed