A Method for the Analysis of Chromium at Sub-μg L<sup>−1</sup> Level Using a Portable Total Reflection X-ray Fluorescence Spectrometer

  • Miyazaki Rihoko
    Department of Industrial Chemistry, Graduate School of Engineering, Tokyo University of Science
  • Oikawa Kosei
    Department of Industrial Chemistry, Faculty of Engineering, Tokyo University of Science
  • Kunimura Shinsuke
    Department of Industrial Chemistry, Faculty of Engineering, Tokyo University of Science Department of Industrial Chemistry, Graduate School of Engineering, Tokyo University of Science

Bibliographic Information

Other Title
  • ポータブル全反射蛍光X線分析装置を用いた水試料中サブμg L<sup>−1</sup>のクロム分析法

Abstract

<p>In this study, total reflection X-ray fluorescence (TXRF) spectra of dry residues of 10 μL, 100 µL, 200 µL, and 400 μL droplets of a solution containing 10 μg L−1 of Cr on hydrophobic film coated sample holders were measured by a portable TXRF spectrometer, and these spectra were compared. The net intensity of the Cr Kα peak per the concentration of Cr in the sample solution (counts / μg L−1) increased with an increase in the volume of a droplet of the sample solution. This was because the mass of Cr in the dry residue increased with an increase in the volume of a droplet of the sample solution. This enhancement in the net intensity of the Cr Kα peak per the concentration of Cr led to a significant improvement in the detection limit expressed as the concentration of Cr. Detection limits for Cr obtained from the measurements of the dry residues of 10 μL, 100 µL, 200 µL, and 400 μL droplets of the sample solution were 2.1 μg L−1, 0.30 μg L−1, 0.13 μg L−1, and 0.09 μg L−1, respectively. Measuring the dry residue of a large volume droplet of a water sample would be useful for the analysis of trace elements in the sample.</p>

Journal

  • Tetsu-to-Hagane

    Tetsu-to-Hagane advpub (0), 2024

    The Iron and Steel Institute of Japan

Details 詳細情報について

Report a problem

Back to top