The Area Measurement Method for Esophageal Lesion under Endoscopy
-
- Narita Ryo
- Akita University
-
- Nansai Shunsuke
- Akita University
-
- Hoshizaki Midori
- Akita University
-
- Manabe Noriaki
- Kawasaki Medical School
-
- Haruma Ken
- Kawasaki Medical School
-
- Oka Kiyoshi
- National Institutes for Quantum and Radiological Science and Technology
-
- Naganawa Akihiro
- Akita University
Bibliographic Information
- Other Title
-
- 内視鏡下における食道病変部の面積計測法
Abstract
The risk of developing esophageal cancer from Barrett’s Esophagus (BE) is associated with its size. However, the study to measure BE area is not advanced well. In this paper, we propose the method to measure BE area from one endoscopic image with camera distance. First, we assume that the esophagus is a cylinder and the camera is a pinhole camera because 3D reconstruction of no-feature points object like an esophagus is difficult. And then, we get BE area indirectly by calculating the surface area of this virtual cylinder. In measurement experiments, we used USB camera and cylinder phantoms to evaluate the accuracy of the proposed method. The errors were about 4-6% in the bottom area, about 10% in the side area, and about 1±2.5 mm in the phantom’s height measurement.
Journal
-
- The Japanese Journal of the Institute of Industrial Applications Engineers
-
The Japanese Journal of the Institute of Industrial Applications Engineers 12 (1), 17-26, 2024
The Institute of Industrial Applications Engineers
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390862422951727360
-
- ISSN
- 21875146
- 2189373X
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
-
- Abstract License Flag
- Disallowed