書誌事項
- タイトル別名
-
- Effects of Iterative Reconstruction on Image Quality of Pediatric Body Computed Tomography Images
- チクジ キンジ オウヨウ サイコウセイホウ ガ CT ガゾウ ノ ブツリ トクセイ ニ オヨボス エイキョウ : ショウニ タイカンブ CT ケンサ オ ソウテイ シタ ケントウ
この論文をさがす
説明
<p>This study evaluated the effects of three types of hybrid iterative reconstruction (IR) on image quality of pediatric body computed tomography images. The image quality components evaluated were noise power spectrum (NPS), task-based modulation transfer function (TTF), and system performance function (SPF). As the IR strength was increased while reducing the radiation dose, the NPS increased in a low-frequency range and the TTF decreased in low-contrast regions. In the low-contrast regions, the calculated SPF decreased over the entire frequency range. Alternatively, in the high-contrast regions, the SPF decreased in the low-frequency regions and increased in the high-frequency regions. The radiation dose reduction using the hybrid IR resulted in the deterioration of the image quality in the low-contrast regions and changes in the spatial frequency characteristics in the high-contrast regions.</p>
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 76 (10), 1035-1043, 2020
公益社団法人 日本放射線技術学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1391975276372775168
-
- NII論文ID
- 130007927931
-
- NII書誌ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 030717624
-
- PubMed
- 33087649
-
- 本文言語コード
- ja
-
- 資料種別
- journal article
-
- データソース種別
-
- JaLC
- NDLサーチ
- Crossref
- PubMed
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可