圧力変換式テンシオメータのキャリブレーション:出力電圧への気温の影響に関する一実験
書誌事項
- タイトル別名
-
- Calibration of Pressure Transducer Tensiometers: A Laboratory Study on Air Temperature Dependence of Pressure Transducer Voltage
- アツリョク ヘンカンシキ テンシオメータ ノ キャリブレーション : シュツリョク デンアツ エ ノ キオン ノ エイキョウ ニ カンスル イチ ジッケン
この論文をさがす
説明
<p>Pressure transducer (pressure sensor) tensiometers are commonly used to monitor the water pressure head in soil. We performed laboratory experiments using a temperature control equipment to clarify the effect of air temperature on output voltage obtained from pressure sensors. We suggested a new calibration formula using the output voltage and air temperature. Under constant pressure head conditions, the decrease in pressure head calculated from the output voltage was linear and ranged between 0.13 and 0.43 cmH2O per 1°C rise in air temperature (4-13 cmH2O per 30°C air temperature rise). The output voltage decreased with an increase in the volume of air in the tensiometers. To obtain the accurate values regarding the pressure head variation, water level in body pipes of the pressure transducer tensiometers should be so high that the part connected to the pressure sensor is immersed in water.</p>
収録刊行物
-
- 地形
-
地形 34 (3), 197-212, 2013-07-25
一般社団法人 日本地形学連合