Aseptic Endoscopy : Application of Newly Developed Endoscope Washer System
-
- FUKUMOTO SHIRO
- Department of Internal Medicine, Shimane Medical University
-
- AMANO YUJI
- Department of Internal Medicine, Shimane Medical University
-
- FUKUDA RYO
- Department of Internal Medicine, Shimane Medical University
-
- ADACHI KYOICHI
- Department of Internal Medicine, Shimane Medical University
-
- GOBARU YOSHIYUKI
- Department of Internal Medicine, Shimane Medical University
-
- ASHIZAWA NOBUO
- Department of Internal Medicine, Shimane Medical University
-
- YOSHIDA HIROSHI
- Department of Internal Medicine, Shimane Medical University
-
- WATANABE MAKOTO
- Department of Internal Medicine, Shimane Medical University
-
- SHIMADA YOSHIHIRO
- Department of Internal Medicine, Shimane Medical University
Search this article
Description
The effect of disinfection on a flexible fiberoptic endoscope with 2% glutaraldehyde solution was compared with disinfection by watery wash alone. The number of bacteria in the biopsy channel decreased from 10^4-10^5 to 10^2-10^3 CFU/ml immediately after watery wash alone. The number, however, increased to 10^5-10^6 CFU/ml after storage in the endoscope cabinet for 3-4 days. The increase in the number of bacteria was caused by proliferation of bacteria such as Pseudomonas aeruginosa, Klebsiella pneumoniae and Proteus mirabilis. The use of these contaminated endoscopes for patients with leukemia or for long term administration of corticosteroid hormone or anticancerous agents was, therefore, concluded to be unsuitable. On the other hand, disinfection with 2% glutaraldehyde solution kept the number of bacteria zero even after storage in the endoscope cabinet for 3-4 days. Endoscope models with perfect waterproof (OLYMPUS OES series) and total immersion endoscope washer have been developed, and total immersion disinfection has made possible. This method of disinfection was easier to operate and more effective than that of conventional disinfection methods, because immersion of the whole endoscope including the operating part became possible.
Journal
-
- Shimane journal of medical science
-
Shimane journal of medical science 10 (2), 72-79, 1986-12-01
Shimane University
- Tweet
Details 詳細情報について
-
- CRID
- 1571980076721006464
-
- NII Article ID
- 110000012481
-
- NII Book ID
- AA00841586
-
- ISSN
- 03865959
-
- Text Lang
- en
-
- Data Source
-
- CiNii Articles