Design and Basic Experiments of "Double Pickup-Coils" Single-Layer High-Tc SQUID Magnetometer
-
- YOKOSAWA Koichi
- Research Institute for Electronic Science, Hokkaido University
-
- AHMAD Rashdi Shah
- Research Institute for Electronic Science, Hokkaido University
-
- MARUYAMA Emi
- Research Institute for Electronic Science, Hokkaido University
-
- KURIKI Shinya
- Research Institute for Electronic Science, Hokkaido University
-
- SUZUKI Daisuke
- Central Research Laboratory, Hitachi Ltd.
-
- KOYANAGI Masao
- Electro-Technical Laboratory
Bibliographic Information
- Other Title
-
- Double Pickup-Coil 型 High-Tc SQUID 磁束計における検出コイル及びSQUID形状の実験的検討
Search this article
Description
A new configuration is proposed for the SQUID and pickup coils of direct-coupling-type high-Tc SQUID magnetometers. The proposed magnetometer has two pickup coils per SQUID, and each pickup coil occupies half the area of the substrate, divided by the bicrystal line. The structure of the pickup coils was optimized by using an analogous model to maximize the effective area of the magnetometer, which is made in a limited substrate area. Longitudinal and transverse type SQUIDs were fabricated with various self-inductances and their performances and coupling coefficients with pickup coils were compared.
Journal
-
- Technical report of IEICE. SCE
-
Technical report of IEICE. SCE 97 (219), 25-30, 1997-07-30
The Institute of Electronics, Information and Communication Engineers
- Tweet
Details 詳細情報について
-
- CRID
- 1571980077315173632
-
- NII Article ID
- 110003187331
-
- NII Book ID
- AN10012885
-
- Text Lang
- ja
-
- Data Source
-
- CiNii Articles