A Modified Wheatstone Bridge for High-Precision Automated Resistance Measurement
-
- Sasaki Hitoshi
- Standards and Measurements Division, Electrotechnical Laboratory
-
- Nishinaka Hidefumi
- Standards and Measurements Division, Electrotechnical Laboratory
-
- Shida Katsunori
- Standards and Measurements Division, Electrotechnical Laboratory
説明
A modified Wheatstone bridge has been developed which auto-balances using a digitally controlled feedback system. The bridge has four detectors to sense the imbalance in the circuit, and a resolution of better than 0.01 ppm with nearly maintenance-free operation.
収録刊行物
-
- Japanese Journal of Applied Physics
-
Japanese Journal of Applied Physics 26 (12), L1947-L1949, 1987