- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Automatic Translation feature is available on CiNii Labs
- Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Clinical Considerations in Medical Imagi Compressed by MPEG-2
-
- NAKAJIMA Isao
- Tokai Univ.
-
- TSURUMI Toyohiko
- Tokai Univ.
-
- SAWADA Yuhwsuke
- Tokai Univ.
-
- MAEDA Toshihide
- Hitachi Ltd.
-
- HAMAMOTO Naokazu
- CTL of MPT
Bibliographic Information
- Other Title
-
- MPEG-2による医療画像評価
Search this article
Description
We mentioned clinical considerations in digital motion pictures(digital fiberscopy, ultrasonic tomogaphy, and injured portion by CCD camera) compressed by MPEG-2. The purpose of this study was to know how much the digital speed might be required to transmite full motion medical pictures on a satellite. The experimental system using CLI MPEG-2 could be changed up to 20 Mbps. The clinical tests were performed by 10 medical doctors at Tokai univ. school of Medicine. The results : ultrasonic tomogaphy might be 1.5 Mbps, digital fiberscopy and injured portion might be 3.0 Mbps.
Journal
-
- IEICE technical report. ME and bio cybernetics
-
IEICE technical report. ME and bio cybernetics 97 (45), 61-66, 1997-05-16
The Institute of Electronics, Information and Communication Engineers
- Tweet
Details 詳細情報について
-
- CRID
- 1572261552315300864
-
- NII Article ID
- 110003286728
-
- NII Book ID
- AN1001320X
-
- Text Lang
- ja
-
- Data Source
-
- CiNii Articles