Effects of blood contact on the properties of tubular dialysis membranes
この論文をさがす
説明
Abstract Dialysis membranes show a decline in hydraulic permeability after blood contact. The objective of the present study is to elucidate the quantitative ch Five kinds of tubular dialysis membranes were employed to determine both urea and pure water permeability before and after blood contact. Constant-pres Pure water permeability decreased after blood contact for all membranes. The drop was marked for membranes of synthetic polymer (PMMA and EVA) but only The experimental results indicate that changes in membrane structure affecting the permeability of dialysis membranes are dependent on pore size and on
収録刊行物
-
- The Chemical Engineering Journal
-
The Chemical Engineering Journal 38 B1-B6, 1988-05-01
Elsevier BV