Monitoring of KrF excimer laser ablation for burn scars: a comparative study of transient reflection measurement and time-resolved photography of ablation plume

この論文をさがす

説明

A simple laser ablation monitoring during burn scar removal by KrF laser irradiation was studied to control laser fluence in real-time. Because, to obtain suitable surface for auto skin-graft, the laser fluence should be precisely controlled at each laser shot. We employed simple probe transmission method which could detect ejected material/phenomena from irradiated surface. The time-course of measured probe intensity contained a couple of attenuated peaks, which might corresponded to a shock wave front and debris plume. The delay time from laser irradiation to the debris plume peak appearance varied with the ablation fluence. The delay time of 1 J/cm 2 (near ablation threshold) case prolonged 25% from 8 J/cm 2 (far above threshold) case. Therefore, we think the delay time measurement by means of the simple probe transmission method may be available to attain the laser fluence control for nonuniform burn scar removal. The time-resolved photography and probe reflection method were also studied to understand the measured time-course of the transmitted probe intensity.

収録刊行物

詳細情報 詳細情報について

  • CRID
    1870865118254284928
  • DOI
    10.1117/12.209897
  • ISSN
    0277786X
  • データソース種別
    • OpenAIRE

問題の指摘

ページトップへ