Application of microcolumnion chromatography using octadecylsilica immobilized with bovine serum albumin as stationary phase for the determination of inorganic anions

この論文をさがす

説明

Microcolumn ion chromatography of inorganic anions has been studied using octadecylsilica immobilized with bovine serum albumin as a stationary phase. The analytes were monitored with a conductivity detector. The detection limits for chloride, nitrate and thiocyanate at S/N = 3 were 0.8, 1.0 and 3.0 μmol L−1, respectively. The present system was applied to the determination of inorganic anions in tap water and serum samples.