Tunable narrow-photon-energy x-ray source using a silicon single crystal

この論文をさがす

説明

A preliminary experiment for producing narrow-photon-energy cone-beam x-rays using a silicon single crystal is described. In order to produce low-photon-energy x-rays, a 100-µm-focus x-ray generator in conjunction with a (111) plane silicon crystal is employed. The x-ray beams from the source are confined by an x-y diaphragm, and monochromatic cone beams are formed by the crystal and three lead plates. The x-ray generator consists of a main controller and a unit with a high-voltage circuit and a 100-µm-focus x-ray tube. In this experiment, the maximum tube voltage and current were 35 kV and 0.50 mA, respectively, and the x-ray intensity of the microfocus generator was 343 μGy/s at 1.0 m from the source with a tube voltage of 30 kV and a current of 0.50 mA. The effective photon energy is determined by Bragg's angle, and the photon-energy width is regulated by the angle delta. Using this generator in conjunction with a computed radiography system, quasi-monochromatic radiography was performed using a cone beam with an effective energy of approximately 15.5 keV.