この論文をさがす
説明
An amorphous silicon photodiode array originally manufactured for facsimile has been applied to an imaging system for proton beam measurements. A Gd/sub 2/O/sub 2/S phosphor sheet and a CsI(Tl) scintillator strip for proton-to-light conversion were tested to obtain good sensitivity. Proton depth dose distribution measurements for cancer therapy are presented. >
収録刊行物
-
- IEEE Transactions on Nuclear Science
-
IEEE Transactions on Nuclear Science 40 569-572, 1993-08-01
Institute of Electrical and Electronics Engineers (IEEE)