Pacing failure caused by automatic pacing threshold adjustment system

DOI PubMed オープンアクセス

この論文をさがす

説明

<jats:title>Abstract</jats:title><jats:p>Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient.</jats:p>