Attenuated total reflectance terahertz spectroscopy for quantitative measurement of somatic cell count in bovine milk
この論文をさがす
説明
Abstract Quantitative somatic cell count (SCC) measurement for detection of mastitis using an attenuated total reflectance (ATR) terahertz (THz) spectroscopy was investigated. Using samples with controlled SCC derived from raw milk, we measured their ATR signal in the THz region 30–440 cm −1 with a Fourier-Transform THz spectrometer. We found ATR signals decreased as SCC increased. Furthermore, we found the correlation between the ATR signal and SCC from 8.0 × 10 4 to 1.57 × 10 6 cells/mL was optimal at 139 cm −1 (R 2 = 098). These findings demonstrate a potential to detect mastitis with a monochromatic THz wave.
収録刊行物
-
- Engineering in Agriculture, Environment and Food
-
Engineering in Agriculture, Environment and Food 8 79-82, 2015-03-01
Asian Agricultural and Biological Engineering Association