Improvement of analysis precision upon the atomic number and electron density measurement by the dual x-ray CT
この論文をさがす
説明
To identify the factor impairing the material identification parameters, which is provided by the dual-energy X-ray computed tomography method using a conventional X-ray tube and a CdTe detector, linear attenuation coefficient was measured by the radioactivity of radio isotopes and compared with theoretical figure. In our study, the atomic number and the electron density is calculated from the linear attenuation coefficient obtained in CT measurement by 64-channel CdTe line detector. To estimate accuracy of CdTe line sensor, it is needed to obtain the linear attenuation coefficient accurately. Using a single detector, the linear attenuation coefficient is verified for accuracy. The energy resolution of CdTe detectors and the method of reconstruction are discussed.