- R. H. Fox
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
-
- A. J. Solman
- Clinical Research Centre, Watford Road, Harrow, Middlesex
-
- R. Isaacs
- Clinical Research Centre, Watford Road, Harrow, Middlesex
-
- A. J. Fry
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
-
- I. C. MacDonald
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
この論文をさがす
説明
<jats:p>1. A new technique for monitoring the deep body temperature is described. The technique depends on creating a zone of zero heat-flow across the body shell; this brings the deep body temperature to the skin surface where it is measured with a simple electronic thermometer.</jats:p> <jats:p>2. The new device gives a temperature closely comparable with other methods for measuring the deep body temperature in the resting subject, and is simple to use and socially acceptable.</jats:p>
収録刊行物
-
- Clinical Science
-
Clinical Science 44 (1), 81-86, 1973-01-01
Portland Press Ltd.