X-ray diffractometer combining synchrotron radiation and pulsed magnetic fields up to 40 T
書誌事項
- 公開日
- 2006-04-13
- 権利情報
-
- http://journals.iucr.org/services/copyrightpolicy.html
- http://journals.iucr.org/services/copyrightpolicy.html#TDM
- DOI
-
- 10.1107/s0909049506006972
- 公開者
- International Union of Crystallography (IUCr)
この論文をさがす
説明
A synchrotron X-ray diffractometer incorporating a pulsed field magnet for high fields up to 40 T has been developed and a detailed description of this instrument is reported. The pulsed field magnet is composed of two coaxial coils with a gap of 3 mm at the mid-plane for passage of the X-rays. The pixel detector PILATUS 100K is used to store the diffracted X-rays. As a test of this instrument, X-ray diffraction by a powder sample of the antiferromagnet CoO is measured below the Néel temperature. A field-dependent lattice distortion of CoO due to magnetostriction is observed up to 38 T.
収録刊行物
-
- Journal of Synchrotron Radiation
-
Journal of Synchrotron Radiation 13 (3), 271-274, 2006-04-13
International Union of Crystallography (IUCr)