還元気化原子吸光法による陸水中のヒ素(III)及びヒ素(V)の分別定量
書誌事項
- タイトル別名
-
- Determination of arsenic(III) and arsenic(V) in fresh water by AAS equipped with arsine generation system.
- カンゲン キカ ゲンシ キュウコウホウ ニ ヨル リクスイチュウ ノ ヒソ 3
この論文をさがす
抄録
For the determination of arsenic(III), arsine was first generated from the sample solution at pH 9 with tetrahydroborate, and then it was introduced into a heated quartz glass cell along with argon; and finally the atomic absorption of arsenic was measured. In addition, the total arsenic was determined under acidic condition (5% hydrochloric acid solution) in the same manner. The arsenic(V) was determined by subtracting the value of arsenic(III) from that of total arsenic. This method was applied for ground and river water samples. Detection levels of 2.6-8.0 ppb of arsenic(V) were detected, whereas arsenic(III) was not detected.
収録刊行物
-
- 分析化学
-
分析化学 40 (4), T97-T100, 1991
公益社団法人 日本分析化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204051114112
-
- NII論文ID
- 110002906263
-
- NII書誌ID
- AN00222633
-
- COI
- 1:CAS:528:DyaK3MXktlGgtrg%3D
-
- NDL書誌ID
- 3713679
-
- ISSN
- 05251931
- http://id.crossref.org/issn/05251931
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- NDL-Digital
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可